At CreativeIT Group we perform leakage current testing for you with the ED&D LT-601 Digital Leakage Current Tester. You can send your equipment to our Monterrey lab, or our engineers can come to your plant, hospital or service center anywhere in Mexico. We run the tests and deliver a report with photo and video evidence and NIST-traceable results. We send you a quote the same day.
What the service covers
Leakage current is a critical parameter in the electrical safety of mains-connected equipment and medical devices. With the LT-601 our engineers evaluate the most relevant configurations:
- Leakage current through the equipment’s ground conductor.
- Leakage current from surface or patient to ground.
- Current between contact points (patient to patient).
The instrument incorporates impedance networks that reproduce the electrical behavior of the human body in accordance with IEC 60601, EN 60601, CSA 601 and UL 2601. It measures on a 200 µA range with 1 µA resolution and a 2000 µA range with 10 µA resolution, with an accuracy of ±1% of full scale. It accepts equipment under test powered at 100–120 or 200–240 VAC, 50–60 Hz, with a maximum current of 15 A.
When you need it
- Electrical safety testing of medical equipment, taking into account IEC 60601 limits for normal and single fault conditions.
- Validation of electronic products before launch or export.
- Preparation for regulatory compliance audits.
- Verification after repair, maintenance or technical service.
- Quality control in the manufacture of electrical and electronic equipment.
If you need a broader evaluation, see our advanced technical evaluation of electrical safety and functional performance service.
How we do it
- We review the type of equipment, its power supply and the reference standard or specification with you.
- We define the measurement configurations and applicable acceptance limits.
- We connect the equipment under test to the LT-601 and measure leakage current in each configuration, under normal condition and, where applicable, under fault condition.
- We document every measurement with photos and video.
- We compare the results against the agreed limits and prepare the report.
What the report contains
- Identification of the equipment under test: brand, model and serial number.
- Measurement configurations and test conditions.
- Measured leakage current values and comparison with the reference limits.
- Photo and video evidence.
- Details of the instrument used, with NIST-traceable results.
The report is available in Spanish and English.
Why CreativeIT Group
- Our own field and lab engineers, with no subcontractors.
- Same-day reply and quote.
- Lab in Monterrey and coverage throughout Mexico.
- We operate the same instrument we offer for rental of the ED&D LT-601.
Frequently asked questions
Can you test medical equipment?
Yes. The LT-601 is compatible with IEC 60601 requirements and includes impedance networks that simulate the human body.
How low are the currents you can measure?
The most sensitive range is 200 µA with 1 µA resolution, suitable for detecting very small leakage currents.
What language is the report in?
We deliver the report in Spanish or English, whichever your company requires.
Request your quote
Tell us which equipment you need evaluated and under which standard, and we will reply with a proposal the same day. Contact us to schedule leakage current testing with our engineers.